Atomic absorption spectroscopy

In analytical chemistry, atomic absorption spectroscopy is a technique used to determine the concentration of a specific metal element in a sample.[1] The technique can be used to analyze the concentration of over 70 different metals in a solution.

Although atomic absorption spectroscopy dates to the nineteenth century, the modern form was largely developed during the 1950s by a team of Australian chemists. They were led by Alan Walsh and worked at the CSIRO (Commonwealth Science and Industry Research Organisation) Division of Chemical Physics in Melbourne, Australia.[2]


Principles

The technique makes use of absorption spectrometry to assess the concentration of an analyte in a sample. It therefore relies heavily on the Beer-Lambert law.

In short, the electrons of the atoms in the atomizer can be promoted to higher orbitals for a short amount of time by absorbing a set quantity of energy (i.e. light of a given wavelength). This amount of energy (or wavelength) is specific to a particular electron transition in a particular element, and in general, each wavelength corresponds to only one element. This gives the technique its elemental selectivity.

As the quantity of energy (the power) put into the flame is known, and the quantity remaining at the other side (at the detector) can be measured, it is possible, from Beer-Lambert law, to calculate how many of these transitions took place, and thus get a signal that is proportional to the concentration of the element being measured.

Instrumentation

Atomic absorption spectrometer block diagram

In order to analyze a sample for its atomic constituents, it has to be atomized. The sample should then be illuminated by light. The light transmitted is finally measured by a detector. In order to reduce the effect of emission from the atomizer (e.g. the black body radiation) or the environment, a spectrometer is normally used between the atomizer and the detector.

Types of atomizer

The technique typically makes use of a flame to atomize the sample,[3] but other atomizers such as a graphite furnace[4] or plasmas, primarily inductively coupled plasmas, are also used.[5]

When a flame is used it is laterally long (usually 10 cm) and not deep. The height of the flame above the burner head can be controlled by adjusting the flow of the fuel mixture. A beam of light passes through this flame at its longest axis (the lateral axis) and hits a detector.

Analysis of liquids

A liquid sample is normally turned into an atomic gas in three steps:

1. Desolvation (Drying) – the liquid solvent is evaporated, and the dry sample remains
2. Vaporization (Ashing) – the solid sample vaporises to a gas
3. Atomization – the compounds making up the sample are broken into free atoms.


Radiation sources

The radiation source chosen has a spectral width narrower than that of the atomic transitions.

Hollow cathode lamps

Hollow cathode lamps are the most common radiation source in atomic absorption spectroscopy. Inside the lamp, filled with argon or neon gas, is a cylindrical metal cathode containing the metal for excitation, and an anode. When a high voltage is applied across the anode and cathode, gas particles are ionized. As voltage is increased, gaseous ions acquire enough energy to eject metal atoms from the cathode. Some of these atoms are in an excited states and emit light with the frequency characteristic to the metal[6]. Many modern hollow cathode lamps are selective for several metals.

Diode lasers

Atomic absorption spectroscopy can also be performed by lasers, primarily diode lasers because of their good properties for laser absorption spectrometry.[7] The technique is then either referred to as diode laser atomic absorption spectrometry (DLAAS or DLAS),[8] or, since wavelength modulation most often is employed, wavelength modulation absorption spectrometry.

Background correction methods

The narrow bandwidth of hollow cathode lamps make spectral overlap rare. That is, it is unlikely that an absorption line from one element will overlap with another. Molecular emission is much broader, so it is more likely that some molecular absorption band will overlap with an atomic line. This can result in artificially high absorption and an improperly high calculation for the concentration in the solution. Three methods are typically used to correct for this:

* Zeeman correction - A magnetic field is used to split the atomic line into two sidebands (see Zeeman effect). These sidebands are close enough to the original wavelength to still overlap with molecular bands, but are far enough not to overlap with the atomic bands. The absorption in the presence and absence of a magnetic field can be compared, the difference being the atomic absorption of interest.

* Smith-Hieftje correction (invented by Stanley B. Smith and Gary M. Hieftje) - The hollow cathode lamp is pulsed with high current, causing a larger atom population and self-absorption during the pulses. This self-absorption causes a broadening of the line and a reduction of the line intensity at the original wavelength.[9]

* Deuterium lamp correction - In this case, a separate source (a deuterium lamp) with broad emission is used to measure the background emission. The use of a separate lamp makes this method the least accurate, but its relative simplicity (and the fact that it is the oldest of the three) makes it the most commonly used method.


Modern developments
Continuum Source AAS

Recent modern developments in electronics and solid state detectors have taken the conventional AAS instrument to the next level. High Resolution Continuum Source AAS (HR-CS AAS) is now available in both flame and graphite furnace mode.

Main features of the new instruments:

* Single xenon arc lamp - Today, multiple hollow cathode lamps are no longer used. With the use of a single xenon arc lamp, all the elements can be measured from 185-900nm. This takes AAS into a true multi-element technique with the analysis of 10 elements per minute.

* CCD technology - For the first time in an AAS CCD chips are now available with 200 pixels which act as independent detectors.

* Simultaneous background correction - Background is now measured simultaneously compared to sequential background on conventional AAS.

* Multiple lines - Extra lines of an analyte are now available thus extending the dynamic working range.

* Better detection limits - Due to the high intensity of the Xenon Lamp there is better signal/noise ratio thus giving better detection limits. In some cases it is up to 10 times better than conventional AAS.

* Direct analysis of solids - In graphite furnace mode it is now possible to analyse solids directly thus avoiding long digestion times.

* Ability to measure sulfur and halogens - It is now possible to measure some non-metals by measuring molecular bands.


Alternatives

For heavy metals such as mercury, alternatives to atomic absorption technology, including direct and pre-concentrated atomic fluorescence techniques such as cold vapour atomic fluorescence spectroscopy can offer higher sensitivity than atomic absorption.[citation needed]

References


1. ^ Sperling, Michael B.; Welz, Bernhard (1999). Atomic Absorption Spectrometry. Weinheim: Wiley-VCH. ISBN 3-527-28571-7.
2. ^ L’vov, B. V. (2005). "Fifty years of atomic absorption spectrometry". Journal of Analytical Chemistry 60: 382. doi:10.1007/s10809-005-0103-0
3. ^ C. T. J. Alkemade, T. Hollander, W. Snelleman and P. J. T. Zeegers, Metal Vapours in Flames, Pergamon Press, Oxford (1982).
4. ^ B. V. L'vov, Forty years of electrothermal atomic absorption spectrometry. Advances and problems in theory, Spectrochim. Acta B 52 1239-1245 (1997).
5. ^ J. A. C. Broekaert, Analytical Atomic Spectrometry with Flames and Plasmas, Third Edition, Wiley-VCH, Weinheim, Germany (1998).
6. ^ Skoog, D.; Holler, J.; Crouch, S. Principles of Instrumental Analysis, 6th ed.; Thomson Books/Cole, 2007; pp 238.
7. ^ O. Axner, Laser Spectrometric Techniques in Analytical Atomic Spectrometry. In: Meyers RA ed. Encyclopedia of Analytical Chemistry, New York, John Wiley & Sons, Inc., (2000), p. 9506-9595.
8. ^ A. Zybin, J. Koch, H. D. Wizemann, J. Franzke and K. Niemax, Diode laser atomic absorption spectrometry, Spectrochimica Acta B 60 1-11 (2005).
9. ^ S. B. Smith, Jr and G. M. Hieftje (1983). "A New Background-correction Method for Atomic Absorption Spectrometry". Applied Spectroscopy 37 (5): 419–424. doi:10.1366/0003702834634893.


See also

* Beer-Lambert law
* Laser absorption spectrometry
* JAAS Journal of Analytical Atomic Spectrometry
* Vapour phase decomposition
* Inductively coupled plasma mass spectrometry